Appendix
IM-Mercury_II_5000_Series Rev. 1 Page 23 Β©2014 MicroE Systems
Mercury II 5000 Series Encoders
Installation Manual and Reference Guide
Maximum Quadrature Output Frequency
Output Frequency (MHz)
A-Quad-B Output Rate
(millions of states/sec)
Dwell Time
(or edge separation) (Β΅sec)
12.50
50.00
0.02
6.25
25.00
0.04
3.125
12.50
0.08
1.563
6.25
0.16
0.781
3.125
0.32
0.391
1.5625
0.64
0.195
0.78125
1.28
0.098
0.390625
2.56
0.049
0.1953125
5.12
0.024
0.09765625
10.24
Note: Values shown are approximate. Exact values may be calculated using either of the
following equations:
𝑂𝑒𝑑𝑝𝑒𝑑 πΉπ‘Ÿπ‘’π‘žπ‘’π‘’π‘›π‘π‘¦ = 12.5𝑀𝐻𝑧/2^𝑛
where n = number of steps below 12.5MHz
𝑂𝑒𝑑𝑝𝑒𝑑 π‘…π‘Žπ‘‘π‘’ = 50/2^𝑛
where n = number of steps below 50 million states per second
6.3 Wiring Diagrams
15-Pin Standard Male D-sub Connector Configuration
Mercury II 15P D-Sub Pin Outs
Quad Output
Pin
Function
1
Right Limit+
2
GND
3
Right Limit-
4
Index-
5
B-
6
A-
7
+5V
8
+5V
9
GND
10
Left Limit+
11
Left Limit-
12
Index+
13
B+
14
A+
15
No Connect
Note: Alarm: A and B are tri-stated if the encoder signal becomes too low for reliable operation.
Note: GND and Inner Shield are internally connected.